X-Rite i1Publish Pro 3 Plus Specifications
SpecificationDetails
ModelX-Rite i1Publish Pro 3 Plus
Experience LevelIntermediate to Advanced
Spectral Enginei1® technology featuring a holographic diffraction grating with 128-pixel diode array and built-in wavelength check.
Spectral Range380 nm to 730 nm
Physical Sampling Interval3.5 nm
Optical Resolution10 nm
Spectral Reporting380 nm to 730 nm in 10 nm increments
Measurement Frequency (Scanning Mode)400 measurements per second
Measurement Geometry45°/0° ring illumination optics, ISO 13655:2017
Measurement Aperture8 mm (0.31") diameter. Effective aperture during scanning depends on patch size and scanning speed.
Illumination Spot Size12 mm (0.47")
Light SourceIntegrated LED illumination including UV
Measurement ConditionsM0 - UV Included (ISO 13655:2017)
M1 - D50 (ISO 13655:2017)
M2 - UV Excluded (ISO 13655:2017)
M3 - Polarization, UV Excluded (ISO 13655:2017)
Optical Brightener CompensationSupported through i1Profiler software
CalibrationManual calibration using external ceramic white reference
Measurement BackgroundWhite, ISO 13655:2017 backup board
Minimum Patch Size (Spot Mode)14 × 14 mm (0.55" × 0.55")
Minimum Patch Size (Scanning Mode)16 × 16 mm with i1iO
16 × 16 mm with sensor ruler
20 × 16 mm without sensor ruler
20 × 16 mm with sensor ruler (M3 Mode)
22 × 16 mm without sensor ruler (M3 Mode)
Maximum Scan Length515 mm (20.28")
Inter-Instrument Agreement0.3 ΔE00 average, 0.8 ΔE00 maximum
Short-Term Repeatability0.05 ΔE00 on white (D50, 2°)
Emission Measurement Range0.2 to 5000 cd/m²
Emission Repeatabilityx,y ±0.002 typical (5000 K, 80 cd/m²)
Ambient Light MeasurementSpectral irradiance (mW/nm/m²), illuminance (lux) with cosine-corrected diffuser head
InterfaceUSB 1.1
Power SupplyPowered directly by USB. No external power supply, charger, or battery required.
Operating Temperature10°C to 35°C (50°F to 95°F)
Operating Humidity0% to 80% non-condensing
Dimensions (Device)162 × 69 × 64 mm (6.4 × 2.7 × 2.5 in.)
Dimensions (Ruler)591 × 122 mm (23.3 × 4.8 in.)
Dimensions (Backup Board)670 × 452 mm (26.4 × 17.8 in.)
Weight285 g (10 oz)

X-Rite i1iO Automatic Scanning Table Specifications
SpecificationDetails
ModelX-Rite i1iO Automatic Scanning Table for i1Pro 3 & i1Pro 3 Plus
Compatible DevicesX-Rite i1Pro 3 and i1Pro 3 Plus Spectrophotometers
Scanning CapabilityYes
Single Pass ScanYes
Scanning SpeedMore than 500 patches per minute in scanning mode under the M0 measurement condition
Measurement BackgroundWhite
Maximum Media Thickness10 mm
Maximum Media Thickness with Optional Z-Axis Spacer33 mm
Minimum Patch Size (Scanning Mode)16 × 16 mm (0.63" × 0.63") when used with the i1Pro 3 Plus
Scanning AreaPositioning area: 32 × 23 cm (12.6" × 9")
Available Disk Space2 GB
CalibrationAutomatic calibration using integrated calibration reference
Communication InterfaceUSB 1.1
ConnectivityPowered USB port
Dimensions (W × D × H)510 × 460 × 170 mm (20 × 18.1 × 6.7 in.)
Experience LevelIntermediate to Advanced
Software ModesBasic mode with step-by-step wizard and Advanced mode for professional color management workflows.
User ExperienceAutomated hands-free operation for fast, accurate, and repeatable color measurements.
Supported LanguagesChinese (Simplified), English, French, German, Italian, Japanese, Korean, Portuguese, and Spanish
Software Development KitAvailable for OEM customers
Windows CompatibilityWindows 7, Windows 8, Windows 8.1, and Windows 10 with the latest Service Pack installed
macOS CompatibilitymacOS X 10.11, 10.12, and 10.13 with the latest updates installed
Unit ColorBlack / Silver
Power Requirements100 to 240 VAC, 50/60 Hz
Warranty12-month limited warranty. Product registration required to receive X-Rite technical support during the warranty period.
Weight4.5 kg (9.9 lb), without i1Pro device installed