Spectral Unit |
Measurement Conditions | According to ISO 13655-2017 Reflection and transmission: M1 (method a), M0, M2 (UVcut), M3 (reflection only, with optional polarization filter) |
---|
Measuring Apertures | Variable between 2, 6 and 8 mm diameter (reflection and transmission) |
---|
Geometry | Reflection 45°:0° circumferential (ISO 5-4:2009); Transmission d:0 (ISO 5-2) |
---|
Calibration | Automatic with internal white reference |
---|
Physical Illumination | Programmable color temperature: Reflection – 3 point circumferential, 7 LED-chip; Transmission – diffuse, 6 LED module |
---|
Measuring Sensor | Diffraction grating with diode array |
---|
Spectral Range | 380 … 750 nm |
---|
Spectral Resolution | 2 nm |
---|
Optical Resolution | 10 nm |
---|
Spectral Reflectance Range | Dynamic (> 200%) for measurement of fluorescent specimens |
---|
Short-Term Repeatability | 0.05 ΔE00 on white ceramic tile (standard deviation, 10 measurements made in spot mode) |
---|
Inter-instrument Agreement | On 12 BCRA color tiles: Average: 0.5 ΔE00, Max: 1.0 ΔE00 |
---|
Air Blowing System | For reflection optics |
---|
Display | 320x240 pixel, touch screen |
---|
Battery | Li-ion accumulator 3,7VDC, 5200mAh |
---|
Charging Time | 2.5 hours when connected to platform, 10 hours through USB port |
---|
Temperature Sensor | Range -20 to +50 °C (-4 to +122 °F); Resolution 0.1, Accuracy: +/- 1.0 |
---|
Humidity Sensor | Range 0 to 100%; Resolution 1, Accuracy: +/- 5% |
---|
Interface | USB 2.0 type C, WiFi 2.4 GHz IEEE 802.11b/g/n (2.4 GHz only); Open, WEP, WPA, WPA2 Security |
---|
Certifications | FCC (USA), IC (Canada), CE (Europe), MIC (Japan), KCC (South Korea), UKCA (UK) |
---|
Dimensions | 237 x 82 x 65 mm |
---|
Weight | 1000 g |
---|